Automatic
measurement on hysteresis loop of soft magnetic material
under dynamic (AC) condition, accurate measurement on
dynamic magnetic characteristic parameters such as
amplitude permeability μa, looss angle δ, iron loss Pc,
remanence Br and coercive force
Hc.
Windows measurement software
applied simply. It conforms to China National Standards
GB3658 - 83, GB5026 - 85 and GB9632 - 88, industry
standard SJ / T10281 - 91 and international standard
IEC60404 - 6.
Analog source (bridge),
frequency indicator, ammeter, voltmeter and wattmeter
are replaced through computer control and high speed A/D
sampling, entire testing process automatically
completed.
Adopt volammetry to measure loss Ps and amplitude
permeability μa and loss angle δ, when test frequency is
50kHz, conduct test on ferrite circular ring sample with
standard size (d1 = 40mm, d2= 32mm, h = 6 ~ 10mm), technical
indices as follows:
Parameters measured
Bm (%)
Br(%)
Hc(%)
μa(%)
Ps(%)
δ(%)
Uncertainty (k=2)
1
1.5
2
3
5
5
Repeatability (constant temperature)
± 0.5
± 1
± 1
± 2
± 3
± 2
Instrument Specifications
WATS - 2010SA Dynamic
Hysteresisgraph
PC5012 A/D Card
Output Power: 200VA sine
wave (below 300kHz) Frequency
Range: 1kHz~ 500kHz Frequency
Fineness: < 0.1% * Current
Value Frequency Error: <
0.05% Output Voltage: 0 ~ 50V
(Peak Value) Voltage Fineness:
Program Control 1mV, panel <
10mV Voltage Distortion
Factor: Superior to
1% Voltage Stability:
Superior to 0.1% Sampling
Current: 0 ~ 5A (Peak
Value) Sample Voltage: 0 ~
10V (Peak Value)