Our company has been engaged in magnetic hysteresis, coercivity, intrinsic coercivity, remanence, magnetic field and other magnetic properties automatic test apparatus development for 20 years. There are many excellent technology professors with profound theory, and strong sense of innovation in our company. Our magnetic hysteresis test products, DMT series hysteresigraph, whose hardware and software are unique in the field, receive lots of home customers' compliments and hold a rather great market control rate, they are ideal for both R&D and production line testing of magnetic materials. When entering into international market, they also win a good reputation. We lay great emphasis on the reliability of quantity and after-sale service, and win universal praise. We heartily appreciate the favor and recommendation of our customers to our products for years and sincerely expect you to become our new customer and new friend

DMT-4 automatic hysteresigraph system accords with the standard permanent magnet test system, GB3217-92, used to measure the major and minor hysteresis loops, demagnetization curves, recoil hysteresis loops and initial magnetization curves in different temperature of ferrite, AlNiCo, SmCo, NdFeB and quenched material, and gives the magnetic parameter, Br, Hcb, Hcj, (BH)max, Hk and so on, and could meet different requirements on test function, precision, power capability, reliability, efficiency, price, local environment from laboratory to enterprise's production line.
The quality of DMT-4 disposes is reliable. There are large-scale, medium-scale and small-scale,7 models in all, of electro-magnets to be chose. Welcome to enquire and ask for data!
Input current:0~30A, total weight is about 350KGs, pole head is changeable, Diameter of pole core is Φ140mm,equipped with wind cooling system. The parameter of magnetic field the generator could generate is as follows:
| Pole max. diameter (mm) | Magnetic field (T) | ||
| Distance 10mm | Distance 30mm | Distance 50mm | |
| Φ110 | 2.43 | 1.75 | 1.35 |
| Φ60 | 3.00 | ||
Input current:0~75A, total weight is about 900KG, pole head is changeable, Diameter of pole core is Φ200mm, equipped with wind and water cooling system, The parameter of magnetic field it could generate is as follows:
| Pole max. diameter (mm) | Magnetic field (T) | |||
| Distance 10mm | Distance 25mm | Distance 35mm | Distance 50mm | |
| Φ110 | 2.90 | 2.537 | 2.33 | 2.06 |
| Φ220 | 2.05 | 1.75 | 1.55 | 1.30 |
Input current:0~200A, the weight of electro-magnet is about 2000KG, pole head is changeable, Diameter of pole core is Φ264mm, equipped with wind and water cooling system, The parameter of magnetic field the power electromagnet could generate is as follows:
| Pole max. diameter (mm) | Magnetic field (T) | |||
| Distance 10mm | Distance 25mm | Distance 35mm | Distance 50mm | |
| Φ110 | 3.25 | 2.85 | 2.62 | 2.35 |
| Φ120 | 3.186 | 2.82 | 2.591 | 2.31 |
| Φ140 | 3.00 | 2.70 | 2.50 | 2.25 |
| Φ220 | 2.60 | 2.30 | 2.13 | 1.91 |
Input current: 0~200A, the weight of electro-magnet is about 2300KG, pole head is changeable, Diameter of pole core is Φ264 mm, equipped with wind and water cooling system, two methods of rise-and-fall, power-operated and hand-operated. The parameter of magnetic field the power electromagnet could generate is as follows:
| Pole max. diameter (mm) |
Magnetic field (T) |
Material | |||
| Distance 10mm | Distance 25mm | Distance 35mm | Distance 50mm | ||
| Φ120 | 3.59 | 3.23 | 3.02 | 2.75 | Pure iron |
| Φ120 | 3.50 | 3.13 | 2.92 | 2.65 | Pure iron |
| Φ120 | 3.38 | 3.04 | 2.84 | 2.60 | Pure iron |
| Φ140 | 3.30 | 3.00 | 2.80 | 2.55 | Pure iron |
* The material could be changed to 2V supermendur (1J220) according to customers' special requirement.
Input current:0~200A, the weight of electro-magnet is about 5000KG, pole head is changeable, diameter of pole core is Φ330mm, equipped with wind and water cooling system, two methods of rise-and-fall, power-operated and hand-operated. The parameter of magnetic field the power electromagnet could generate is as follows:
| Pole max. diameter (mm) \ Current (A) |
Magnetic field (T) | |||
| Distance 10mm | Distance 15mm | Distance 25mm | Distance 35mm | |
| Φ120 (200A) | 3.48 | 3.26 | 2.95 | 2.69 |
| Φ110 (200A) | 2.3 | 2.1 | 2 | 1.7 |